Calcium measurement using ion selective electrode based on a benzo-18-crown-6 as an ionophore | ||
| Journal of Chemical Reactivity and Synthesis | ||
| دوره 10، شماره 4، اسفند 2020، صفحه 175-185 اصل مقاله (361.68 K) | ||
| نوع مقاله: Research Article | ||
| نویسندگان | ||
| Shaghayegh Lovineh1؛ Arezoo Ghaemi* 2 | ||
| 1Department of Chemistry, khouzestan Science and Research Branch, Islamic Azad University, Ahvaz, Iran | ||
| 2Department of Chemistry, Ahvaz Branch, Islamic Azad University, Ahvaz, Iran | ||
| چکیده | ||
| A new PVC based membrane using benzo-18-crown-6 as an ionophore has been developed as a calcium ion selective electrode. The electrode comprises of ionophore (4%), PVC (30%) and plasticizer (63%). The electrode shows a linear dynamic response in the concentration range of 1×10−5 M to 1×10−1 M with a Nernstian slope of 29.8 mV/decade and detection limit is 4.0×10−6 M. The proposed sensor shows reasonably good selectivity with respect to alkali, alkaline earth and transition metal ions. It can be used in the pH range of 5 to 12. The proposed electrode also was applied for the direct determination of Ca2+ cation in doubly distilled deionized water samples spiked with different concentrations of calcium. This electrode was also used for determination of Ca2+ in real samples with a very good recovery. | ||
| کلیدواژهها | ||
| Calcium (II) cation؛ Ion selective electrode؛ Potentiometry | ||
|